Color touchscreen LCD display, Program. Management Onboard™, Real CPR Help® and CPR
Dashboard. Configurable to also display ECG in addition to CPR Dashboard.
Protocol: Semiautomatic and Fully Automatic configuration
Waveform: ZOLL Rectilinear Biphasic™
Defibrillator Charge Hold Time: 30 seconds; ZOLL AED 3 Automatic model: 3 seconds prior to automatic shock delivery
Energy Selection: Factory preprogrammed selection (Adult: 120 J, 150 J, 200 J; Child: 50 J, 70 J, 85 J)
Patient Safety: All patient connections are electrically isolated
Charge Time: Less than 10 seconds with new battery
Pre-shock Pause: 8 seconds with new battery
Electrodes: ZOLL CPR Uni-padz™
Self-test: Configurable automatic self-test every day or every 7 days. Default: every 7 days. Monthly full-energy test (200 J).
Automatic Self-test Checks: Battery capacity, status and expiration; electrode connection and expiration; ECG and charge/ discharge circuits; microprocessor hardware and software; CPR circuitry and pads sensor; audio circuitry
Compression Depth: 1.9 cm to 10.2 cm; 0.75 in to 4 in (adult patients only)
Defibrillation Advisory: Evaluates defibrillation pad connection and patient ECG to determine if defibrillation is required.
Shockable Rhythms: Ventricular fibrillation with average amplitude >100 microvolts and wide complex ventricular tachycardia with rates greater than 150 BPM for adults, 200 BPM for pediatrics. For ECG analysis algorithm sensitivity and specificity, refer to AED 3 Administrator’s Guide.
Patient Impedance Measurement Range: 10 to 300 oh
Defibrillator: Protected ECG circuitry
Display Format: High-resolution LCD with capacitive touch panel
Display Screen Size: 5.39 cm x 9.5 cm; 2.12 in x 3.74 in
Data Recording and Storage: User-configurable for 1 or 2 clinical events for total of 120 minutes. Includes ECG, impedance measurements, device prompts and CPR data.
Data Recovery: Controlled by touchscreen, uploaded to USB memory stick, or ZOLL Case Review™ over a Wi-Fi network
Internal Clock Synchronization: Coordinated Universal Time (UTC) synchronization occurs when communicating with the ZOLL Online server.
Size: (H x W x D) 12.7 cm x 23.6 cm x 24.7 cm; 5.0 in x 9.3 in x 9.7 in
Weight: 2.5 kg; 5.5 lbs (with battery)
Power: Battery
Wireless: 802.11 a/b/g/n
Security Protocols: WPA 1, WPA 2, WPA Personal, WPA Enterprise
Port: USB 2.0
Rescue Accessory Pack Contents: Non-latex gloves, a CPR barrier mask, razor, paper towel, and towelette.
Device Classification: Class II and internally powered per EN60601-1
Design Standards: Meets applicable requirements of EN 60601-1, IEC 60601-1-11, IEC 60601-2-4
Operating Temperature: 0° to 50°C; 32° to 122°F
Storage Temperature: -30° to 70°C; -22° to 158°F
Humidity: 10% to 95% relative humidity, non-condensing
Vibration: IEC 60068-2-64, Random, Spectrum A.4, Table A.8, Cat 3b; RTCA/DO-160G, Fixed Wing Aircraft, Section 8.6, Test Cat. H, Aircraft Zone 1 and 2, EN 1789, Sweep per EN 60068-2-6 Test Fc.
Shock: IEC 60068-2-27; 100G
Altitude: -381 m to 4573 m; -1250 to 15,000 ft
Particle and Water Ingress: IP55
Drop Test: 1 metre; 3.28 feet
Type: Disposable, sealed lithium manganese dioxide
Battery Standby Life (once installed): 5 years with weekly self-test. End of life indicated by blank status window (typical remaining shocks: 9).
Battery Shelf Life: Store for up to 2 years at 23°C (77°F) prior to installing in AED 3 to maintain above battery life.
Temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 10% to 95% (non-condensing)
Weight: 317.5 grams; 0.7 lbs.
Size: (H x W x D) 27.75 mm x 133 mm x 88 mm; 1.0 in x 5.16 in x 3.5 in
Nominal Voltage: 12 volts
Shelf Life: 5 years
Conductive Gel: Polymer Hydrogel
Conductive Element: Tin
Packaging: Multilayer foil laminate pouch
Impedance Class: Low
Cable Length: 142 (+/-3.8) cm; 56 (+/- 1.5) in
Design Standards: Meets requirements of IEC 60601-2-4, 2010