Broad Featured Capability: Invasive and non-invasive support for infant (5 kg and up) to adult patients, with a flexible array of ventilation options including Pressure Control/Volume Control ventilation, SIMV/Assist Control spontaneous breath support, and stand-alone CPAP/Bilevel support.
Patient Range: Infant/Pediatric (5 kg) to Adult
Controls/Settings
Modes
Tidal Volume: 40 to 2,000 ml
RR (Respiratory Rate): 0 or 5 to 60 b/min
I Time (Inspiratory Time): 0.5 to 2.0 sec
PEEP/CPAP: 0 to 25 cmH2O/0 to 25 mbar
PS (Pressure Support)**: 0 to 25 cmH2O / 0 to 25 mbar*
*PEEP + PS cannot exceed 25 cmH2O/mbar Operating Humidity 5 to 95% non-condensing
Flow: 5 to 60 L/min (flows > 80 LPM in PS Mode, CPAP and demand flows)
Pressure Control (Target Pressure): 15 to 55 cmH2O / 15 to 55 mbar
Ptrig (Sensitivity): -5 to -1 cmH2O / -5 to -1 mbar
Manual Inflation: 2 second maximum
FiO2: .21 to 1.00
Maximum I Time: 2 seconds
Flow Wave Pattern: Square or descending**
** Based on mode selected
Auto Lock Function: ON / OFF
Remote Alarm Option: Normally Open/Closed
Monitored Parameters
Paw (Peak, Base)
Delivered Tidal Volume
Spontaneous Breath Count
Integrated Battery Level
External Battery Level – When Attached
Calculated Parameters
I:E Ratio
Flow rate
FiO2
Alarms
High Airway Pressure: 10 to 55 cmH2O/10 to 55 mbar
Low Airway Pressure: 5 to 15 cmH2O/5 to 15 mbar
Back-up Ventilation Alert: 20 second fixed, apnea ventilation mode
Low Source O2: Triggers at 40-37 psi
Low Source Air: Triggers at 40-37 psi
Device Alert: Malfunction alert
Critical Battery: Approx. 20 minutes of battery charge remains
Silence/Reset: 110 sec
Pneumatics
Gas Delivery System: Internal dual diaphragm
Air/Oxygen Mixer: Integrated
Oxygen Input: 41 to 87 psig oxygen
Power Requirements
AC Voltage: 110- 240 VAC, 50/60 Hz, max. 1A
DC Voltage: 11-15 VDC, max. 5A
Integrated Battery Parameters
Battery Composition: Sealed Lead Acid
Run time; 100% O2: 7.5 hours
Run time; 60% O2: 7.5 hours
Run time; 100% air (on compressor): 7 hours
W/ External Battery Pack Parameters (MCV-AUXBAT)
Battery Composition: Sealed Lead Acid
Run time; 100% O2: 21 hours
Run time; 60% O2: 21 hours
Run time; 100% air (on compressor): 21 hours
Dimensions & Weight
Width: 13.3″ (34 cm)
Depth: 5.3″ (13 cm)
Height: 13.8″ (35 cm)
Weight: 18.8 lbs. (8.5 kg)
Environmental
Shock: 100 G-Force
Vibration: To IEC 60068-2-6
Operating Temperature: 0° F to 122° F (-18° C to 50° C)
Storage Temperature: -40° F to 140° F (-40° C to 60° C)
Operating Humidity: 5 to 95% non-condensing
Water Ingress: IPx 22
Signal Connections
Remote Alarm: 1/4″ phone jack to connect to nurse call or remote system
Specialty Features
Quick Start: Programmable presets to quickly recall pre-determined ventilation protocols
Custom Modes: 3 additional programmable custom modes that are available for quick recall
Oxygen Conserve: Allows the use of up to 10 LPM of oxygen at 3 to 40 psi through the DISS port or up to 10 LPM of oxygen at less than 10 psi through the low pressure barb fitting for use with oxygen concentrator